Toyo Corporation*, Japan
MIA-5M Material Impedance Analyzer
MIA-5M Material Impedance Analyzer
The Material Impedance Analyzer MIA-5M can be used in a variety of applications such as complexes, toners, and organic semiconductors.
- High resistance of up to 1 TΩ.
- 0.2% high precision measurement with full open/short/load correction function.
- The calibration function enables high accuracy measurement in a short time without the need for a measurement reference. Especially in the high frequency range of 100 kHz and above.
- Fully automatic operation, centralized control of high pressure amplifier and temperature by control software.
[System Introduction]
When a material needs to be industrialized, we must grasp most of its properties as widely as possible and with high precision, so as not to affect the direction of industrialization at a later stage due to wrong data and cause unnecessary losses.
The MIA series is an AC impedance evaluation system for various dielectric materials developed jointly by Toyo Teknika Co. The system is equipped with an exclusive OSL compensation system, achieving a measurement error accuracy of up to 0.2%.
[System Features]
Corresponding to samples up to 1 TΩ - Suitable for dielectric materials, coordination compounds, ionic conductive materials, color toners, organic semiconductors, etc.
High accuracy - base measurement accuracy 0.2% - with OSL correction
Dedicated software - HIMS fully automated control measurement software - Impedance analysis software such as Z-View
Various specialized accessories - Clamping options for different material forms and sizes - Options for various gas atmospheres, environments and temperatures
[Optional Accessories]
Voltage amplification function - up to DC: ±4kV, AC: 800Vpp, with special protection device
Sample fixture/temperature control
Model | TTPX | LN-Z2 | Constant temperature box+SH2-Z | HT-22-1000 | UHT-22-1200 |
Name | Cryogenic Probe Station | Wide Area Temperature Console | Heat resistant sample table | Desktop Oven | Ultra High Temperature Oven |
Temperature control range | 4.2K ~ 425K | 80K ~ 473K | -60℃ ~ +150℃ | 100℃ ~ 1,000℃ | 100℃ ~ 1,200℃ |
*Depends on model refrigerant | |||||
Sample contact method | Probes | Upper and lower clips | Upper and lower electrodes | Upper and lower clips | Upper and lower clips |
Atmospheric gas control | Vacuum, gas replacement | Vacuum, gas replacement | Different sizes correspond to different specifications | Flow rate control | Flow rate control |
[System Specifications]
Parameters | |
---|---|
Sample contact method | 2 end contact |
Frequency range | 1mHz ~5MHz |
Basic measurement accuracy | 0.20% |
Impedance measurement range | 100Ω ~ 1TΩ |
Capacitance measurement range | 1pF ~ 1mF |
Dielectric loss (tanδ) measurement range | 0.0001 ~ 1000 |
Applied voltage (AC + DC Vpp) | -10V ~ +10V |
Correction function | OSL |
Control Software | Comes with HIMS software, PC control |
Related Applications